Non-contact temperature measurement starts with a simple question: do you need a value at one point, or the temperature distribution across a surface? A pyrometer measures infrared radiation from a single spot and turns it into a process signal. A thermal imaging camera does the same for tens of thousands of points at once. The choice depends on the material, the temperature range, the target geometry and the integration into your automation system.
Key takeaways
- Pyrometer: a control value at a well-defined point, fast response, a 4–20 mA signal straight into the PLC.
- Thermal imaging camera: when you do not know in advance where the hot spot will appear, when the product moves or when uniformity matters.
- Choose the wavelength by material: short wavelengths for metals, dedicated bands for glass and plastic films, 8–14 μm for most non-metals.
- Ratio pyrometers reduce the influence of variable emissivity, dust and dirty windows, but do not remove it completely.
- The distance-to-spot ratio (D:S) must be checked at the real mounting distance.
Spot or area measurement
A pyrometer integrates radiation from a spot whose size depends on the optics and the distance. If the process has a clear control point (a billet leaving the induction coil, the strip entering the rolling mill, the glass gob), a fixed pyrometer gives a stable value that is easy to integrate. Response can be very fast: the IMPAC IS 6 Advanced has a t90 of 120 μs, and the IMPAC Highspeed 740 series goes down to 6 μs.
A thermal imaging camera sees a matrix of pixels, each with its own temperature. It is the right tool when the hot spot changes position, when you track a profile across the width of the product, or when one installation replaces several measuring points. The Optris Xi 80, Xi 410, Xi 1M and Xi 05M cameras can run stand-alone: they find the hot spot and output the value on an analogue or alarm output, without a PC. For moving products, the PIX Connect software of the Optris PI cameras has a line-scan function.
Often the two complement each other: the camera locates problems, the pyrometer closes the control loop. Browse our pyrometers and thermal imaging cameras, including portable versions for maintenance.
Wavelength: choose by material
Emissivity depends on the material and on the wavelength. For metals, the rule is to work at the shortest wavelength compatible with the temperature range: there, the same emissivity error produces a much smaller temperature error. The price is the lower limit: silicon detectors (approx. 0.7–1.1 μm) usually measure from 550 … 700 °C, InGaAs detectors (1.45–1.8 μm) from lower temperatures, and extended InGaAs versions (2–2.6 μm) from 75 … 100 °C.
Glass is transparent in the near infrared; around 5 μm it becomes opaque and the surface is measured. Thin flat glass below 1 mm requires 7.8–7.9 μm. Thin plastic films are measured in their absorption bands, at 3.43 μm or 7.9 μm.
| Material / application | Typical wavelength | Examples from our range |
|---|---|---|
| Metals at high temperatures | 0.8–1.1 μm (Si) | IMPAC IS 6 Advanced, Optris CT 1M, Xi 1M, PI 1M |
| Metals from medium temperatures | 1.45–1.8 μm; 2–2.6 μm (InGaAs) | IMPAC IGA 6 Advanced, IGA 320/23, Optris CT 2M / 3M |
| Molten metal | 0.5–0.55 μm | Optris Xi 05M, PI 05M, CTlaser 05M; IMPAC IS 8-GS Pro |
| Glass surface | ≈ 5 μm | IMPAC IN 5/5 plus (5.14 μm), Optris CT G5 |
| Thin flat glass | 7.8–7.9 μm | IMPAC IN 6/78, Optris PI 450i G7 / PI 640i G7 |
| Thin plastic films | 3.43 μm or 7.9 μm | Optris CT P3 / CT P7 |
| Non-metals, painted metals | 8–14 μm | Optris CS LT, CT LT, Xi 410, PI 640i LT; IMPAC IN 5 |
| Through flames / flue gas | 3.9 μm; 4.24 μm (CO₂); 4.64 μm (CO) | Optris Xi 320 MT; Optris CTlaser F2 / F6 |
Emissivity: the main source of error
Non-metals (ceramics, paint, rubber, food, thick plastics) have a high, stable emissivity in the 8–14 μm band. Shiny metals have a low emissivity that changes with oxidation, surface finish and temperature. A low-emissivity surface also reflects radiation from its surroundings: the walls of a hot furnace can push the reading up.
Set the emissivity by comparison with a contact thermocouple, avoid reflections, and when measuring through a window enter its transmission as well. The instrument itself is checked separately, against a traceable black-body source; the Mikron sources in our calibration range cover both pyrometers and thermal imaging cameras.
Ratio (two-colour) pyrometers
A ratio pyrometer measures simultaneously at two neighbouring wavelengths and calculates the temperature from the ratio of the signals. If emissivity drops equally in both channels, or if dust, steam or a dirty window attenuate the signal in the same proportion, the ratio stays practically unchanged. According to the manufacturer, the IMPAC IGAR 6 Smart also measures targets smaller than the spot and signals a dirty window on a relay (“Dirty Window”).
The limits: materials whose emissivity changes differently in the two channels need a slope correction (the Optris CTratio and CSvision provide a Smart Ratio Mode for this), and the lower limit is higher (on the IGAR 6 Smart, 250 °C in ratio mode versus 100 °C in one-colour mode). Options in our range: the Optris CTratio, with fibre optics, up to 3000 °C; the Optris CSvision, with a video camera, up to 3500 °C; the IMPAC IGAR and ISR series.
Distance-to-spot ratio and optics
At a D:S ratio of 100:1 and 1 m distance, the spot is about 10 mm. The target must fill the spot completely, with a margin for vibration and alignment; otherwise a one-colour pyrometer averages the object with the cooler background. Focusable optics give the smallest spot at the focus distance: the IMPAC IS 6 Advanced has about 350:1, with a 0.6 mm spot at 210 mm. For small targets, sighting matters: dual laser (Optris CTlaser), through-the-lens or video (Optris CSvideo, CTvideo).
For cameras, the equivalent is the field of view of one pixel. As a practical rule, the object should cover several pixels in each direction. The Optris Xi 320 MT, for example, reaches D:S 550:1 with the 11° lens.
Outputs, interfaces and mounting
Pyrometers offer a two-wire 4–20 mA output (IMPAC IN 5, IS 50-LO/GL), 0–10 V or a J/K thermocouple output (Optris CT), plus RS485, Modbus RTU, Profibus DP, Profinet or Ethernet, depending on the model. Cameras communicate over Ethernet/PoE or USB; the Optris Xi accepts process interfaces with up to nine analogue or alarm outputs. The environment often decides the model: the Optris CT LT head runs without cooling up to 180 °C ambient, the IMPAC IS 50-LO/GL fibre up to 250 °C; beyond that, cooled housings and air purging are used.
Frequently asked questions
Can I measure steel with an 8–14 μm pyrometer?
Only on heavily oxidised or painted surfaces with stable emissivity. On shiny steel or steel with variable oxidation the error can be large; for metals choose a short-wavelength pyrometer (Si or InGaAs).
When is one camera better than several pyrometers?
When there are many points, when their position varies, or when you need the profile across the width. A camera with measuring areas defined in software and process interfaces can replace several fixed pyrometers.
Does a ratio pyrometer completely remove the influence of emissivity?
No. It compensates for equal changes at both wavelengths and for uniform signal attenuation. If the emissivity ratio changes, a slope correction is needed, determined by comparison with a reference.
Send us the material, temperature range, mounting distance, target size, response time and the signal your automation system expects. We will recommend the right pyrometer or camera and prepare an offer: request a quote.