Optris CTlaser — industrial pyrometers with dual-laser sighting
Dual-laser infrared pyrometers with separate measuring head and electronics, up to 2200 °C, with variants for non-metals, metals, glass, plastic and combustion gases.
| Temperature range | CTlaser LT: −50 … 975 °C; 05M: 1000 … 2000 °C; F2/F6: 200 … 1450 °C; F2H/F6H: 400 … 1650 °C; up to 2200 °C depending on variant |
|---|---|
| Spectral range | 8–14 µm (LT); 525 nm (05M); 4.24 µm (F2, CO₂); 4.64 µm (F6, CO) |
| Optical resolution (D:S) | 75:1 (LT); 50:1 (LT F); 150:1 (05M); 45:1 (F2/F6) |
| Minimum spot size | 0.9 mm at 70 mm (LT, CF1 lens); 1.6 mm at 70 mm (F2/F6) |
| Response / exposure time (90 %) | 120 ms (LT); 9 ms (LT F); 1 ms (05M); 10 ms (F2/F6) |
- A quote tailored to your application
- Technical advice on choosing the right model
- Technical support before and after delivery
Description
Optris CTlaser pyrometers use dual-laser sighting: the two beams mark the actual position and size of the measuring spot at any distance. The target can be positioned exactly even when it is small or far away. The measuring head is separate from the electronics and operates without cooling at ambient temperatures up to 85 °C. The laser switches off automatically at 50 °C. Depending on the variant, the range extends up to 2200 °C.
Key benefits
- Optics with selectable focus: spots from 0.9 mm (CTlaser LT) and an optical ratio of up to 150:1 (CTlaser 05M).
- Fast variants: 9 ms on the CTlaser F (LT) and 1 ms exposure time on the CTlaser 05M.
- 0/4–20 mA, 0–5/10 V, J/K thermocouple and alarm outputs. Optional USB, RS232, RS485, CAN, Profibus DP, Ethernet interfaces.
- Accessories for harsh environments: cooling housing, water cooling and air purge.
Typical use
- CTlaser LT / CTlaser F: non-metals, plastic, fast-moving objects at low temperatures (−50 … 975 °C).
- CTlaser 05M / 1M / 2M / 3M / 4M: metals and metallurgy; 05M for molten metal (1000 … 2000 °C).
- CTlaser F2 / F6: CO₂ or CO combustion gases, 200 … 1650 °C, in burners, incinerators and chemical reactors.
- CTlaser G5 / G7 / P7: glass, thin glass sheets and plastic films.
Applications
- Metallurgy and molten metal
- Glass industry
- Plastics processing
- Combustion and incineration processes
- Chemical reactors
- Small or fast-moving objects
Technical specifications
| Temperature range | CTlaser LT: −50 … 975 °C; 05M: 1000 … 2000 °C; F2/F6: 200 … 1450 °C; F2H/F6H: 400 … 1650 °C; up to 2200 °C depending on variant |
|---|---|
| Spectral range | 8–14 µm (LT); 525 nm (05M); 4.24 µm (F2, CO₂); 4.64 µm (F6, CO) |
| Optical resolution (D:S) | 75:1 (LT); 50:1 (LT F); 150:1 (05M); 45:1 (F2/F6) |
| Minimum spot size | 0.9 mm at 70 mm (LT, CF1 lens); 1.6 mm at 70 mm (F2/F6) |
| Response / exposure time (90 %) | 120 ms (LT); 9 ms (LT F); 1 ms (05M); 10 ms (F2/F6) |
| Accuracy | ±1 % or ±1 °C (LT); ±1 % of reading up to 1100 °C (05M) |
| Analog outputs | 0/4–20 mA, 0–5/10 V, J/K thermocouple |
| Alarm output | Open collector, 24 V / 50 mA; optional 2 optically isolated relays |
| Digital interfaces (optional) | USB, RS232, RS485, CAN, Profibus DP, Ethernet |
| Power supply | 8–36 V DC |
| Ambient temperature | Head and electronics: −20 … 85 °C (head: 50 °C with the laser on) |
| Sighting laser | 635 nm, 1 mW, switched on/off from the electronics or software |
| Protection | IP65 (NEMA-4) |
| Weight | Head: 600 g; electronics: 420 g |
Documentation
Request a quote for this product
Fill in the form; a TempSystem engineer will reply with a quote and, if needed, with questions about your application.
- Send the request (the product is filled in automatically).
- We check the application and the right variant.
- You receive the technical and commercial offer.