Optris CSlaser 2M — dual-laser pyrometer for metals
Two-wire infrared pyrometer for metals, 250 … 1600 °C, 1.6 µm, D:S up to 300:1, 10 ms response time, with dual-laser sighting.
| Temperature range | 250 … 800 °C (2ML); 385 … 1600 °C (2MH); scalable in software |
|---|---|
| Spectral range | 1.6 µm |
| Optical resolution (D:S) | 150:1 (2ML); 300:1 (2MH); minimum spot 0.5 mm (2MH, CF2 lens at 150 mm) |
| Accuracy | ±(0.3 % of reading + 2 °C) |
| Repeatability | ±(0.1 % of reading + 1 °C) |
- A quote tailored to your application
- Technical advice on choosing the right model
- Technical support before and after delivery
Description
The Optris CSlaser 2M is a compact two-wire infrared pyrometer for measuring the temperature of metals, parts from secondary metal processing and ceramic materials. It operates at 1.6 µm, between 250 °C and 1600 °C. Measuring spots start at 0.5 mm. Dual-laser sighting marks the target exactly, which matters for small parts or long distances.
Key benefits
- Standard two-wire 4–20 mA interface, for simple integration into a PLC.
- Optical ratio of 150:1 (2ML) or 300:1 (2MH) and 10 ms response time.
- Fast parameterization and real-time measurement over USB, with the Compact Connect software.
- Operates without cooling at ambient temperatures up to 85 °C (50 °C with the laser on).
- IP65 housing that can be front-mounted in vacuum processes, down to 10⁻³ mbar.
Typical use
Temperature measurement of metals and ceramic materials, secondary metal processing and vacuum processes, with direct PLC integration.
Applications
- Metal measurement
- Secondary metal processing
- Ceramic materials
- Vacuum processes
Technical specifications
| Temperature range | 250 … 800 °C (2ML); 385 … 1600 °C (2MH); scalable in software |
|---|---|
| Spectral range | 1.6 µm |
| Optical resolution (D:S) | 150:1 (2ML); 300:1 (2MH); minimum spot 0.5 mm (2MH, CF2 lens at 150 mm) |
| Accuracy | ±(0.3 % of reading + 2 °C) |
| Repeatability | ±(0.1 % of reading + 1 °C) |
| Response time (90 %) | 10 ms |
| Analog output | 4–20 mA, two-wire (max. loop resistance 1000 Ω) |
| Alarm output | Open collector, 0–30 V / 500 mA |
| Digital interface | 9.6 kBaud; optional USB |
| Power supply | 5–30 V DC |
| Ambient temperature | −20 … 85 °C (50 °C with the laser on) |
| Protection | IP65 (NEMA-4); front mounting in vacuum down to 10⁻³ mbar |
| Weight | 600 g |
Documentation
Request a quote for this product
Fill in the form; a TempSystem engineer will reply with a quote and, if needed, with questions about your application.
- Send the request (the product is filled in automatically).
- We check the application and the right variant.
- You receive the technical and commercial offer.